SDG1000 User Manual 56
N-Cycle/Gated
N-Cycle has specific number of waveform cycles, and every burst is activated
by a trigger event. Gated burst use external source to control burst as when
to be activated.
Figure 2- 47 Table 2- 24 Menu Explanations of the N-Cycle Parameters (Page2/2)
Cycles
Set the number of waveform cycle in an N-Cycle (1 to 50,000 or Infinite).
If you choose Infinite, then a continuous waveform will be generated which
will not stop until a trigger event happens.
If needed, Burst Period will increase to cater to the specific number of
cycles.
For an infinite-cycle burst, external or manual trigger is needed to
activate burst.
Function
Menu
Settings
Explanation
Trig Out
Off
Signal Triggered at Rise Edge.
Signal Triggered at Fall Edge.
Turn off Trigger Setting.
Cycles/
Infinite
Set the number of the bursts in a N-Cycle.
Set the number of the bursts in a N-Cycle to
be infinite.
Delay
Set the delay time before the burst starts.