MK LIT (E5) 01/2021 page 96/172
“Ramp” and “Step” mode example (representation of a temperature course)
W/°C
t/min.
0
20
40
60
80
100
1
0
30 40 60 70 100
130
1 2 3 4 5 6 7 8
Corresponding program table
Section No.
Duration
[hh:mm:ss]
Temperature
[°C]
Ramp or Step
1
00:10:00
40.0
Step
2
00:20:00
60.0
Step
3
00:10:00
80.0
Step
4
00:20:00
40.0
Step
5
00:10:00
40.0
Ramp
6
00:30:00
80.0
Ramp
7
00:30:00
80.0
Ramp
8
00:00:01
20.0
Ramp
17.7.3 Special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line Idle modeactivates / deactivates the operating mode Idle mode”.
Operation lines Switching output 1“ up to „4“ can be used to turn on and off any equipment connected
to the zero-voltage switching outputs (DIN sockets (17) and (18) (option, chap. 25).
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air connec-
tion (for optional compressed air connection, chap. 26.5)
Operation line Anti-condensation protectionserves to switch the anti-condensation protection (chap.
24).
The other operation lines are without function.