Binder LIT MK 240 dynamic climate chambers

User Manual - Page 83

For LIT MK 240. Also, The document are for others Binder models: LIT MK 720

PDF File Manual, 173 pages, Read Online | Download pdf file

LIT MK 240 photo
Loading ...
Loading ...
Loading ...
MK LIT (E5) 01/2021 page 83/172
15.3.3 Message and measures in the state of alarm
The state of alarm is indicated visually in Normal display by the alarm message “Safety controller alarm
and the screen header flashing in red color. If the buzzer is enabled (chap. 13.2.5) there is an additional
audible alert (chap. 13.2.3). The alarm remains active until it is acknowledged on the controller and the
inner temperature falls below the set safety controller setpoint. Then the heating is released again.
Normal display with safety controller alarm.
Press the Alarm icon.
List of active alarms.
Press the Reset alarm icon.
15.3.4 Function check
Check the safety controller at appropriate intervals for its functionality. It is recommended that the author-
ized operating personnel should perform such a check, e.g., before starting a longer work procedure.