MK LIT (E5) 01/2021 page 123/172
The waiting icon shows how much time there is left to touch the currently activated box. If
the box is not touched within this period, calibration is aborted and the display changes to
Normal display.
After completing the calibration, i.e., touching all four boxes, the display changes to Normal display.
20.5 Network and communication
For these settings at least the “Admin” authorization level is required.
20.5.1 Serial interfaces
The chamber is optionally equipped with a serial RS485 interface.
This menu allows to configure the communication parameters of the RS485 interface.
The device address is required to recognize chambers with this interface type in a network, e.g. when
connecting it to the optional BINDER APT-COM™ 4 Multi Management Software (chap. 26.1). In this case
do not change the other parameters.
Path: Main menu > Settings > Serial interfaces
“Serial interfaces” submenu.
Select the desired setting in the field Baud rate”.
Select the desired setting in the field Data format”.
Select the field Minimum response timeand enter the desired minimum response time. Confirm entry
with Confirm icon.
Select the field Device addressand enter the device address. Factory setting is “1”. Confirm entry with
Confirm icon.
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.