MK LIT (E5) 01/2021 page 140/172
Data Logger Kit T 220: Temperature range -90 °C / -130 °F up to +220 °C / 428 °F
For detailed information on installation and operation of the BINDER Data Logger, please refer
to the mounting instructions Art. No. 7001-0204 and to the original user manual of the manu-
facturer, supplied with the data logger.
26.4 Analog outputs for temperature (option)
With this option, the chamber is equipped with analog outputs 4-20 mA for actual value and set-point value
of temperature. These outputs allow transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (19) in the lateral control panel as follows:
ANALOG OUTPUTS TEMPERATURE 4-20 mA DC
PIN 1: Temperature actual value
PIN 2: Temperature actual value +
PIN 4: Temperature set-point value
PIN 5: Temperature set-point value +
Temperature range: -40 °C / -40 °F up to +180 °C / -356 °F
A suitable DIN plug is enclosed.
Figure 25: Pin allocation of the DIN socket (19) for option analog outputs
26.5 Object temperature display with flexible Pt 100 temperature sensor (option)
The object temperature display enables the determination of the actual temperature of the charging material
during the whole process. The object temperature is measured via a flexible Pt100 temperature sensor and
can be viewed on the controller display. The sensor top protective tube of the flexible Pt 100 can be im-
mersed into liquid substances.
Normal display with object temperature display
(sample values)
The object temperature data is given out together with the data of the temperature controller to the Ethernet
interface as a second measuring channel and can be documented by the APT-COM™ 4 Multi Management
Software (option, chap. 26.1) developed by BINDER.
Technical data of the Pt 100 sensor:
Three-wire technique
Class B (DIN EN 60751)
Temperature range up to 320 °C / 608 °F
Stainless steel protective tube 45 mm / 1.78 in length, material no. 1.4501