MK LIT (E5) 01/2021 page 12/172
The chambers were produced in accordance with VDE regulations and were routinely tested in accordance
to VDE 0411-1 (IEC 61010-1).
A function test is carried out on each safety module before delivery.
To operate the chamber, use only original BINDER accessories or accessories from third-party suppliers
authorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories.
1.7.1 Notes on the installation site
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the chamber.
Do NOT install the chamber in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Observe the prescribed minimum distances when installing the chamber (chap. 8.4)
DANGER
Electrical hazard by water entering the chamber.
Deadly electric shock.
The chamber must NOT become wet during operation, cleaning, or maintenance.
Do NOT install the chamber in damp areas or in puddles.
Set up the chamber in a splash-proof manner.
Do not install or operate the chamber in hazardous locations.
DANGER
Danger of explosion due to combustible dusts or explosive mixtures in the vicinity
of the chamber.
Serious injury or death from burns and / or explosion pressure.
Do NOT operate the chamber in potentially explosive areas.
KEEP combustible dust or air-solvent mixtures AWAY from the chamber.
WARNING
Danger of injury and damages by the chamber tipping over or breakaway of the pro-
truding lower housing cover.
Injuries and damage to the chamber and the loading material
Do NOT load the lower housing cover with heavy objects while the chamber door is
open and do NOT climb on it.