MK LIT (E5) 01/2021 page 66/172
Place the CO
2
pressurized gas cylinder in the bracket provided and secure it with the strap.
Securing the CO
2
pressurized gas cylinder with
the strap
Inserted and fixed CO
2
pressurized gas cylinder
11.1.2 Activating the gas detection system
Switch on the activation switch of the gas detection system.
Attention! When the gas detection system is activated via the main switch, the CO
2
fire suppression
device may be triggered automatically. In that case, CO
2
flows quickly from the pressurized gas cylinder
into the interior of the chamber. If the door is open, there is a risk of suffocation and frostbite.
DANGER
Risk of suffocation, poisoning, and frostbite due to triggering of the CO
2
fire
suppression device.
Death due to suffocation. Frostbite.
Do NOT open the chamber door when the gas detection system is active.
Switch the main switch of the gas detection system off before opening the door of
the chamber.
After switching the gas detection system on, wait approx. 5 minutes until the sensors are initialized
(during initialization, sensors show "Err 2")
In the meantime, check the flow meter for measuring gas and dilution gas and adjust, if necessary.
IMPORTANT: The measuring and dilution gas must always be set to a ratio of 1:1, otherwise the CO
and H
2
sensors will malfunction. A deviation from the volumetric flow set point will result in a fault notifi-
cation and activation of the yellow indicator light (flashing signal).
After successful initialization, the display changes to the measured gas concentration