MK LIT (E5) 01/2021 page 141/172
26.6 Compressed air connection (option)
This option permits directly connecting compressed air to the chamber.
Requirements for the compressed air supplied directly to the chamber
Quality of the air: DIN ISO 8573-1:2010 [2:2:1]
Supply pressure: 6-8 bar domestic connection
If a different connection pressure is required, please contact BINDER INDIVIDUAL Customized Solu-
tions.
Permissible temperature: +10 °C / 50 °F up to +50 °C / 122 °F.
Air requirement: 15 m³/h (at normal pressure)
Connection:
Connection is established to the coupling connector (25) in the rear panel : Standard quick acting closure
socket for compressed air, nominal width 7.85 mm / 0.31 in.
Activation:
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air connection
(25).
27. Cleaning and decontamination
Clean the chamber after each use in order to prevent potential corrosion damage by ingredients of the
loading material.
Prior to renewed startup, allow the chamber to completely dry after all cleaning and decontamination
measures.
The safety module must NOT be opened for cleaning purposes or poured over with water.
DANGER
Electrical hazard by water entering the chamber and the safety module.
Deadly electric shock. Damage of components of the safety equipment.
Do NOT spill water or cleaning agents over the inner and outer chamber surfaces.
Do NOT spill water or cleaning agents over the safety module.
Do NOT open the safety module for cleaning.
Do NOT put ANY cleaning aids (cloth or brush) into slots or openings on the chamber.
Before cleaning turn off the main switch of the gas detection system.
Before cleaning, turn off the chamber at the main power switch and disconnect the
power plug. Let the chamber cool down to ambient temperature.
Completely dry the chamber before turning it on again.