MK LIT (E5) 01/2021 page 52/172
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the voltage and frequency changer.
Do NOT install the voltage and frequency changer in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
The voltage and frequency changer is equipped with four castors. The rear castors can be easily locked
via the attached brakes
9.4 Electrical connection
9.4.1 Information on connecting the battery test chamber
The chambers are supplied ready for connection.
The chambers come with a fixed power connection cable of at least 1800 mm / 70.87 in in length. They are
equipped with three internal overload releases against excess-current.
Model Power plug
Nominal voltage +/- 10% at the
indicated power frequency
Current type Chamber fuse
LIT MK 115
LIT MK 240
LIT MK 720
CEE plug 5-poles
16 Amp
400 V at 50 Hz 3 N~
16 Amp
3 x internal
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chambers protective conductor meets the latest
technology. The protective conductors of the socket and plug must be compatible!
DANGER
Electrical hazard due to missing protective conductor connection.
Deadly electric shock.
Make sure that the chamber’s power plug and the power socket match and se-
curely connect the electrical protective conductors of the chamber and the house
installation.
Only use original connection cables from BINDER according to the above specification.
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chambers type plate (left chamber side, bottom right-hand, see chap. 1.6)
NOTICE
Danger of incorrect power supply voltage due to improper connection.
Damage to the chamber.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
When connecting, please observe the regulations specified by the local electricity supply company as
well as the local or national electrical regulations (VDE directives for Germany).
Observe a sufficient current protection according to the number of devices that you want to operate. We
recommend the use of a residual current circuit breaker.