Automated Pressure Calibrator
Troubleshooting
87
Troubleshooting
Table 17 explains minor troubleshooting issues. For issues outside of the scope of this
section, the Product may require service. See Contact Fluke Calibration.
Table 17. Troubleshooting
Problem Probable Cause Action
Electrical Problems
Product does
not turn on
Not plugged in Verify Product is plugged in and power is available.
Power not available Check the fuse per instruction in this manual.
Fuse blown
Verify the main power switch is ON.
Check the fan. If the fan is on or solenoid click is heard, see
the “Display does not turn on” problem.
If the fan is not on, have the power supply serviced. Verify
internal power connections.
Display does
not turn on
Chassis power issue Verify that power is supplied, see above.
Screen saver is
activated
Verify power to front panel. and should be lit.
Power plug to back
panel is loose
Verify the is in the ON state.
Display panel failure
Open front panel and verify the power plug is connected to
the front panel.
Electrical Problems
Rear panel
drivers not
operating
Loose connection Inspect the connection.
Exceeded maximum
power rating
Allow internal fuse to cool and retry. Verify the rating of the
external solenoids.
Limit current of external solenoids.
EMM Function Problems
No electrical
measurement
is shown
EMM not installed Install EMM (see Module Installation)
Wrong connection
between UUT and
Product
Connect UUT to Product with correct “+” and “-“ terminal.
Measurement over
range
Make sure the input current of voltage is within the
measurement range (0 mA-24 mA, 0 V-30 V).