Automated Pressure Calibrator
Calibration Tasks
53
UUT
To configure the pressure device under test, push and enter edit mode. The
following parameters are available.
Use UUT Tolerance: Yes/No
Select yes or no to use the UUT tolerance or not.
Threshold Tolerance:
Setting threshold tolerance value
Relative Tolerance:
Setting relative tolerance value
Combination Method:
Great of /Addition
Combine the Threshold Tolerance and Relative
Tolerance. At each pressure you would calculate the
threshold and the relative and use whichever is greater
or as addition.
New
Use New to make a new pressure device program.
Push to select and name a new program. Put the correct value fill in the
column:
Start: The start pressure in the test sequence
End: The end pressure in the test sequence
Points Up: Determine the numbers of upscale.
Points Down: Determine the numbers of downscale.
Dwell Time:
See “Dwell Time” under the Transmitter section.
Max Time:
See “Max Time” under the Transmitter section.
Auto Fill:
See “Auto Fill” under the Transmitter section.
Delete
Select Delete to remove the selected program name. Push to select and
delete.
Rename
Select Rename to change the name of a selected program. Push to select
and Rename.
Push to save the new program name.