2271A
Operators Manual
74
Test Port Insert
For devices with 1/8 BSP or NPT mounting threads, the diameter of the thread is
very close to the effective sealing diameter of the O-ring fitted to the test port.
See Table 15 and Figure 22.
This can make it difficult to achieve a good seal. When mounting these devices,
use the test port insert (stored in the spare seals container).
1
2
3
4
hwr038.eps
Figure 22. Test Port Insert
Table 15. Test Port Inserts - Parts List
Item Description Part
1 O-ring 3865163
2 Test Port Insert 3919892
3 O-ring 3865195
4 Test Port 4542465
To calibrate panel-mounted gauges with pressure connections in the rear, use an
Angle Adapter such as the Fluke P5543. The Product operates the CPS without
any further interaction required by the operator.