Automated Pressure Calibrator
External Driver Communication
77
External Driver Communication
The Product has four external 24 V dc powered drivers. These drivers support
several optional accessories and can support the driving of solenoids for
customized solutions. Use the display and remote interface to set driver status.
Each driver is a low-side switched 24 V dc component. Each Channel includes a
PTC-type resettable fuse.
Maximum continuous power per channel is 10 W and total continuous power
maximum for all 4 channels combined is 24 W. See Figure 24.
1
0
0
-
2
4
0
VAC
4
7
-
6
3
Hz
1
0
0
W
MA
X
T2A
2
5
0
V
S
UP
PLY
M
WP: 23 MPa
EXHAUST
VENT
M
WP: 150
k
P
a
FLUKE
COR
P
ORATION
w
ww
.fluk
e
c
a
l.com
SERIAL NU
MBER
R
E
F
E
REN
CE
P
ORT
M
W
P
: 1
5
0
kPa
Drivers
hwr008.eps
Figure 24. Drivers
Configure the Drivers
Turn on the CPS and/or isolation valve accessories from the Instrument Setup
menu (>Instrument Setup). Both require the correct pneumatic
connections and electrical connections. Select the External 24V tab to see or set
the state of the external drivers.
Drivers are identified in the External 24V menu and on the rear panel of the
Product as DRV1, DRV2, DRV3, DRV4 (see Figure 25). Each solenoid uses two
wires, one wire goes into each connector.
DRV1 uses the two top left inputs
DRV2 uses the two top right inputs
DRV3 uses the two bottom left inputs
DRV4 uses the two bottom right inputs