2271A
Operators Manual
70
CPS Use
To use the CPS, connect the UUT to the top test port on the CPS using the same
method as described in the Connect a UUT section.
WCaution
To prevent Product damage or damage to a gauge:
DO NOT use PTFE tape on these connections. This prevents
a correct seal. The gauge adapter sealing system can be
hand-tight sealed up to 20 MPa (3000 psi). Wrenches or
similar tools are not required. Over tightening can cause
damage to threads or sealing faces.
Before connection, ensure that there is an O-ring fitted to
the test port.
Check that the sealing face of the device to be fitted is clean
and undamaged, as scratches or dents can form leak-paths.
Note
The threads on the test port, and the lower part of the gauge
adapters are left-handed.
1. Screw on the appropriate gauge adapter to the UUT, see Figure 18.
hwr034.eps
Figure 18. Screw on Gauge Adapter