Operators Manual
48
Exercise
The Exercise function repeatedly pressurizes a UUT to reduce the effects of
hysteresis. The Exercise menu specifies a maximum and minimum Setpoint
pressure and how many times to repeat the routine (cycles).
To Exercise, set these parameters so the Product can do the task:
Maximum – Max Setpoint
Minimum – Minimum Setpoint
Cycle – How many times to repeat the tests
Delay – How long to stay at setpoint when the setpoints are reached
Control at Target – If ON, the Product uses the dynamic control mode to keep
pressure at the Setpoint. If OFF, the Product uses the static pressure control at
the Setpoint.
Zero PMM
Use this function to manually set the PMM zero-pressure value to be the same
as the local atmosphere pressure value.
Note
Zeroing is done automatically for HART tasks when in gauge mode.
CPS Cleanout (If CPS is installed)
Use the CPS Cleanout menu to choose the duration of the CPS cleaning
process. The process removes contaminates from the CPS.
To run a cleanout process:
1. Select and enter the CPS cleanout menu.
2. Set the cleanout time in seconds.
3. Push to run the process, push to repeat the cleanout cycle if
needed.
4. Push to stop.