2271A
Operators Manual
62
Hart Diagnostics
Some HART devices feature a self-test function that check various aspects of the
transmitter such as the processor and memory when commanded. Not all HART
devices have this feature. If the transmitter does have this function, the Product
sends the self-test command and shows the results of the self-test on the
display. If the feature is not implemented in the transmitter, no information will is
shown.
Start HART Diagnostics test:
1. Select (HART Tasks) on the HART Menu.
2. Select HART Diagnostics.
3. Push to start.
4. HART Diagnostics Results dialog is shown. Errors or faults (if any) are shown
if occurred.
hwr012.jpg
Write PV Unit
Changes the Primary Variable (pressure unit) of the transmitter.
Note
This can also be changed in the HART Data menu.
1. Push (HART Tasks) on the HART menu.
2. Select Write PV Unit and push to open the PV unit menu.
3. Select the new unit.
4. Push (Send).
5. Push to exit.