190 Series III
Users Manual
82
To turn on the display again:
Press any key. The Display will be visible again and the Display Auto-Off timer starts again.
The display will be turned off again when the time has elapsed.
Connect the power adapter and the Auto-Off timer is inactive.
Auto Set Options
With the next procedure you can choose how auto set behaves when you press the AUTO-
MANUAL (auto set) key.
1. Press
U to show the USER key labels.
2. Press 1 to open the USER OPTIONS menu.
3. Use the cursor and E to open the AUTO SET ADJUST menu.
If the frequency range is set to >15 Hz, the Connect-and-View function responds more
quickly. The response is quicker because the test tool is instructed not to analyze low
frequency signal components. However, when you measure frequencies lower than 15 Hz,
the test tool must be instructed to analyze low frequency components for automatic
triggering:
4. Use the cursor and E to select 1 Hz and up, then jump to Input Coupling.
When you press the AUTO-MANUAL (auto set) key, the input coupling can either be set to dc
or left unchanged.
5. Use the cursor and E to select Unchanged.
When you press the AUTO-MANUAL (auto set) key glitch capture can either be set to On or
left unchanged.
6. Use the cursor and E to select Unchanged.
Note
The auto set option for the signal frequency is similar to the automatic trigger option for
the signal frequency. See Automatic Trigger Options. However, the auto set option
determines the behavior of the auto set function and shows only effect when you press
the auto set key.