190 Series III
Users Manual
28
Figure 13. Resistor Value Readings
Current Measurement
You can measure current in both Scope mode and Meter mode. Scope mode has the advantage
of waveforms being displayed while you perform measurements. Meter mode has the advantage
of high measurement resolution.
The next example explains a typical current measurement in Meter mode.
XW Warning
Carefully read the instructions about the current probe you are using.
To set up the test tool:
1. Connect a current probe, for example, Fluke i410 (optional) between the 4-mm banana jack
inputs and the conductor to be measured.
2. Ensure that the red and black connectors correspond to the red and black banana jack inputs.
See Figure 14.
3. Press
M to show the METER key labels.
Figure 14. Measurement Setup
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com