FLUKE MDA-550/FPC Motor Drive Analyzer with one-year Premium Care, 500 MHz, 4-channel

User Manual for the Fluke MDA-550/FPC Motor Drive Analyzer with one-year Premium Care, 500 MHz, 4-channel - Page 79

For MDA-550/FPC.

PDF File Manual, 97 pages, Read Online | Download pdf file

MDA-550/FPC photo
Loading ...
Loading ...
Loading ...
190 Series III
Users Manual
74
Independently Floating Isolated Inputs
The test tool has independently floating isolated inputs. You can use the independently floating
isolated inputs to measure signals that are independently floating from each other. Independently
floating isolated inputs offer additional safety and measurement capabilities compared to inputs
with common references or grounds.
Each input section (A, B, C, D, A, B, METER INPUT) has its own signal input and its own
reference input. The reference input of each input section is electrically isolated from the
reference inputs of the other input sections. The isolated input architecture makes the test tool
about as versatile as having four independent instruments. The advantages of having
independently floating isolated inputs are:
Simultaneous measurement of independently floating signals.
Additional safety:
Since the commons are not directly connected, the chance of causing short circuit when
measuring multiple signals is greatly reduced.
When measuring in systems with multiple grounds, the ground currents induced are kept
to a minimum.
Because the references are not connected together inside the test tool, each reference of the
used inputs must be connected to a reference voltage. Independently floating isolated inputs are
still coupled by parasitic capacitance. This can occur between the input references and the
environment, and between the input references mutually. See Figure 38. For this reason, you
should connect the references to a system ground or another stable voltage. If the reference of
an input is connected to a high speed and / or high voltage signal, you should be aware of
parasitic capacitance. See Figure 38, Figure 40, Figure 41, and Figure 42.
Note
The input channels are electrically isolated from the USB port and from the power
adapter input.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...