Replay, Zoom and Cursors
39
Cursor Measurements
Cursors allow you to make precise digital measurements on waveforms. This can be done on live
waveforms, recorded waveforms, and on saved waveforms.
Horizontal Cursors on a Waveform
To use the cursors for a voltage measurement:
1. From Scope mode, press
G to show the cursor key labels.
2. Press 1 to highlight the horizontal cursor icon.
3. Press 2 to highlight the upper cursor icon.
4. Use
h i
to move the upper cursor position on the screen.
5. Press 2 to highlight the lower cursor.
6. Use
h i
to move the lower cursor position on the screen.
Note
Even when the key labels are not displayed at the bottom of the screen, you can use the
cursor. This allows full control of both cursors while having full screen view.
The screen shows the voltage difference between the two cursors and the voltage at the cursors.
See Figure 22. Use horizontal cursors to measure the amplitude, high or low value, or overshoot
of a waveform.
Figure 22. Voltage Measurement with Cursors