190 Series III
Users Manual
16
Smart Average
In the normal average mode occasional deviations in a waveform just distort the averaged wave
shape, and do not show up on screen clearly. When a signal really changes, for instance when
you probe around, it takes quite some time before the new wave shape is stable. With smart
averaging you can quickly probe around, and incidental waveform changes like a line flyback in
video show up on screen instantly.
Figure 7. Smoothing a Waveform
Persistence, Envelope and Dot-Join to View Waveforms
You can use Persistence to observe dynamic signals. See Figure 8.
1. Press
P to show the SCOPE key labels.
2. Open the WAVEFORM OPTIONS menu with 4.
3. Use the cursor and E to highlight to Waveform: and open the Persistence... menu.
4. Use the cursor and E to select:
a. Digital Persistence: Short, Medium, Long or Infinite to observe dynamic waveforms like on
an analog oscilloscope.
b. Digital Persistence: Off, Display: Envelope to see the upper and lower boundaries of
dynamic waveforms (envelope mode).
c. Display Dot-join: Off to show measured samples only. Dot join off may be useful when
measuring for example modulated signals or video signals.
d. Display: Normal to turn the envelope mode off and the dot-join function on.