Manual - Page 186

For SDS5032X.

Loading ...
Loading ...
Loading ...
SDS5000X Series Digital Oscilloscope User Manual
186 / 236 W W W. S I G L E N T. C O M
single point outside the mask will cause a failure.
All Out: All data points must be outside the mask to pass the test. Even
a single point inside the mask will cause a failure.
Any In: Any data point inside the mask will be recognized as pass. All
data points outside the mask will cause a failure.
Any Out: Any data point outside the mask will be recognized as pass.
All data points inside the mask will cause a failure.
23.4 Operation
Touch Operation to start/stop the test. Stopping a test in progress and restart
the test will clear the count of the passed frames, failed frames, total frames,
and the fail rate. Pressing the Clear Sweeps button on the front panel can also
clear the count information.