16.5.2 CAN Trigger
Touch Trigger Setting in the
CAN trigger dialog box to
set the trigger condition:
Start The oscilloscope triggers at the beginning of the frame.
Remote The oscilloscope triggers on a remote frame with a specified
ID. ID, ID Bits (11-bit or 29-bit) and Curr ID Byte (1st, 2nd, 3rd, or 4th byte)
can be set. Curr ID Byte is used to specify the byte to be adjusted when
using the universal knob.
ID The oscilloscope triggers on the data frame that matches the
specified ID. ID, ID Bits (11-bit or 29-bit), and Curr ID Byte (1st, 2nd, 3rd, or
4th byte) can be set.
ID+Data The oscilloscope triggers on the data frame that matches the
specified ID and data. ID, ID Bits (11-bit or 29-bit), Curr ID Byte (1st, 2nd,
3rd, or 4th byte), Data1 and Data2 can be set.
Error The oscilloscope triggers on the error frame.
16.5.3 CAN Serial Decode
The set of SPI decode is similar to that of I2C decoding.