SIGLENT
SSA3000X User Manual 63
2.6 Shortcut Key
2.6.1 Preset
Recall the preset setting and restore the analyzer to a specified status.
Press System ->Pwr On/Preset ->Preset to select “Def”, “Last” or “User”.
Press Preset to load the factory settings listed in the following table (except items marked with **”)
or User-defined settings.
Table 2-40 Factory Settings
Parameter
Default
Frequency
Center Freq
1.6 GHz
Start Freq
0 Hz
Stop Freq
3.2 GHz
Freq Step
Auto
Span
Span
3.2 GHz
X Scale\
Linear
Amplitude
Ref Level
0 dBm
Attenuator
Auto, 20 dB
Preamp
Off
Units
dBm
Scale/Div
10 dB
Scale Type
Log
Ref Offset
0 dBm
Corrections
Off
BW
RBW
Auto, 1 MHz
VBW
Auto, 1 MHz
VBW/RBW
1
Avg Type
Log Pwr
Filter
Gauss
Sweep
Sweep Time
Auto
Sweep Rule
Speed
Sweep
Continue
Sweep Mode
Auto
Numbers
1
Trig