Siglent REFL-SSA3000X Reflect Measurement Software for SSA3000X Series Spectrum Analyzers

Manual - Page 54

For REFL-SSA3000X.

PDF File Manual, 96 pages, Read Online | Download pdf file

REFL-SSA3000X photo
Loading ...
Loading ...
Loading ...
SIGLENT
SSA3000X User Manual 36
Default
0 dB
Range
-20 dBm ~ 0 dBm
Unit
dBm
Knob Step
1 dB
Direction Key Step
10 dB
2.2.7.3 TG Level Offset
Assign a certain offset to the output power of the TG when gains or losses occur between the TG
output and external device in order to display the actual power value.
This parameter only changes the readout of the TG output power, rather than the actual value.
The offset could be either a positive (gain in the external output) or a negative (loss in the external
output).
Table 2-22 TG Level Offset
Parameter
Explanation
Default
0 dB
Range
-200 dB ~ 200 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
10 dB
2.2.7.4 Normalize
Normalization can eliminate the error of TG Level. Before using this function, connect the [TG
SOURCE] output terminal of the TG with the [RF INPUT] input terminal of the analyzer.
When enabled, the reference trace will be stored automatically after the current sweep finishes if no
reference trace is stored before. During the reference trace storage, the corresponding prompt
message is displayed.
When normalization is enabled, the corresponding value of the reference trace will be subtracted from
the trace data after every sweep.
Default
reference plane
Normalized
reference plane
Loading ...
Loading ...
Loading ...