SSA3000X User Manual 38
demodulations are available in this device.
2.2.8.1 Demod (AM/FM)
Set the demodulation type to AM or FM; or disable the demodulation function. The default is off.
The system will enable a marker automatically, place it at the center frequency and perform AM (or
FM) demodulation on this frequency point after you enable AM (or FM) demodulation.
Analyzer provides earphone jack and the demodulated signal can be output in audio frequency
(AF) mode through the earphone. The frequency and intensity of AF denotes the frequency and
amplitude of the signal respectively.
2.2.8.2 Earphone
Set the status of the earphone. When it is on, the demodulated signal can be heard through the
earphone during the demodulation. By default, it is off.
2.2.8.3 Volume
Set the volume of the earphone.
Table 2-25 Volume
Parameter
Explanation
Default
6
Range
0 ~ 10
Unit
N/A
Knob Step
1
Direction Key Step
1
2.2.8.4 Demod Time
Set the time for the analyzer to complete a signal demodulation after each sweep. If Earphone is set to
“On”, you will hear the demodulated signal through the earphone during the demodulation. A longer
demod dwell time would be better to demodulate audio signal.
Table 2-26 Demod time
Parameter
Explanation
Default
5 s
Range
5 ms ~ 1000 s
Unit
ks, s, ms
Knob Step
0 ms ~ 100 ms, step=1 ms
100 ms ~ 1 s, step =10 ms