Siglent REFL-SSA3000X Reflect Measurement Software for SSA3000X Series Spectrum Analyzers

Manual - Page 50

For REFL-SSA3000X.

PDF File Manual, 96 pages, Read Online | Download pdf file

REFL-SSA3000X photo
Loading ...
Loading ...
Loading ...
SIGLENT
SSA3000X User Manual 32
2. Numbers
Set the number of sweeps for a single sweep. In single sweep mode, the system executes the specified
number of sweeps and the number shown on the icon in the status bar at the left of the screen varies
with the process of the sweep.
3. Continue
Set the sweep mode to “Continue”. The character Cont on the parameter icon denotes the
analyzer is sweeping continuously.
If the instrument is in single sweep mode and no measurement function is enabled, press this key
and the system will enter continuous sweep mode and sweep continuously if the trigger conditions
are satisfied.
If the instrument is in single sweep mode and a measurement function is on, press this key and the
system will enter continuous sweep mode and measure continuously if the trigger conditions are
satisfied.
In continuous sweep mode, the system will send a trigger initialization signal automatically and
enter the trigger condition judgment directly after each sweep.
Table 2-16 Sweep Times
Parameter
Explanation
Default
1
Range
1 ~ 9999
Unit
N/A
Knob Step
1
Direction Key Step
1
2.2.4.4 Sweep Mode
Sweep mode includes auto, sweep and FFT.
1. Auto
When the sweep mode is auto, the analyzer selects the sweep mode automatically between
Sweep and FFT Mode in the shortest time.
2. Sweep
Work in point-by-point scanning. The Sweep mode is only available when RBW is in 30 Hz 1 MHz
3. FFT
Work in FFT scanning. The FFT mode is only available when RBW is in 1 Hz - 30 kHz.
When TG is on, the sweep mode is forced to Sweep.