Siglent REFL-SSA3000X Reflect Measurement Software for SSA3000X Series Spectrum Analyzers

Manual - Page 71

For REFL-SSA3000X.

PDF File Manual, 96 pages, Read Online | Download pdf file

REFL-SSA3000X photo
Loading ...
Loading ...
Loading ...
SIGLENT
SSA3000X User Manual 53
Default
2 MHz
Range
100 Hz ~ Sweep Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Integration BW /100, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
3. Adjacent channel bandwidth
Set the frequency width of the adjacent channels.
The adjacent channel bandwidth is related to the main channel bandwidth and the range available is
from main channel bandwidth/20 to main channel bandwidth×20.
Table 2-34 adjacent channel bandwidth
Parameter
Explanation
Default
2 MHz
Range
100 Hz ~ Sweep Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Integration BW /100, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
4. Adjacent Channel space
Set the difference between the center frequency of the main channel and the center frequency of the
adjacent channels.
Adjusting this parameter will also adjust the distance between the upper/lower channel and the main
channel.
Table 2-35 Adjacent channel space
Parameter
Explanation
Default
2 MHz
Range
33 Hz ~ full span
Unit
GHz, MHz, kHz, Hz
Knob Step
Integration BW /100, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
Loading ...