SSA3000X User Manual 12
3, Load the .lic file provided by pressing File ->“Load from internal memory or USB stick.
The option will be enabled after rebooting.
1.7.3 Firmware Upgrade
Follow this procedure to update the instrument firmware:
1. Download the firmware package from an official SIGLENT website. www.siglentamerica.com,
http://www.siglent.com/ens, https://www.siglenteu.com
2. Extract and copy the .ADS file into the root directory of a USB stick.
3. Plug the USB stick into the USB Host connector. Press System->System Info-> Firmware
Update”; find the .ADS file in USB stick.
4. Press the Load, the analyzer will perform the update process automatically.
Remarks:
The upgrade process will take several minutes. When the upgrade is completed, the machine will
reboot.
Any interruption during the update process will result in update failure and system data loss. This
is not covered under the warranty and the user will bear repair costs and shipping.
Do not remove the USB storage device until the update is finished.
1.8 Menu Operation
There are 7 types of menus according to their operation modes. Each type of menu and its operation
method are introduced below.
1. Parameter Input
When selected, use the numeric keys to modify the parameters.
For example, select Center Freq, input the desired figure and press Enter
to change the center frequency.
2. State switching
Press the corresponding menu key to switch between the sub-options.
For example, press Preamp to enable or disable the amplifier.
3. Enter Lower Menu(with parameter)
Press the corresponding menu key to enter the lower menu and change the
option currently selected.
For example, press Units to enter the lower menu. Select dBm and return
to the previous menu. The unit of Y-axis changes to dBm.
4. Enter Lower Menu (without parameter)
Press the corresponding menu key to enter the lower menu.
For example, press Calibration to enter the lower menu directly.
5. Direct Execution