SSA3000X User Manual 43
measurement values of multiple points. The table allows for up to eight markers to be displayed at one
time.
Figure 2-8 Marker table
2.3.2 Marker ->
1. M->CF
Set the center frequency of the analyzer to the frequency of the current marker.
◆ If Normal marker is selected, the center frequency will be set to the frequency of the current
marker.
◆ If Delta or Delta Pair marker is selected, the center frequency will be set to the frequency of the
Delta Marker.
◆ The function is invalid in Zero span mode.
2. M -> CF Step
Set the center frequency step of the analyzer to the frequency of the current marker.
◆ If Normal marker is selected, the center frequency step will be set to the frequency of the current
marker.
◆ If Delta or Delta Pair marker is selected, the center frequency step will be set to the frequency of
the Delta Marker.
◆ The function is invalid in Zero span mode.
3. M -> Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.