Binder FDL 115 Safety drying chambers, series FDL – 115-liter capacity

User Manual - Page 62

For FDL 115.

PDF File Manual, 77 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
FDL (E2.1) 02/2019 page 62/76
17.5 Equipment and Options (extract)
To operate the safety drying oven, use only original BINDER accessories or accessories /
components from third-party suppliers authorized by BINDER. The user is responsible for any
risk arising from using unauthorized accessories.
Regular equipment
Electronically controlled APT.lineheating technology
In compliance with all safety requirements according to EN 1539: 2015, EN ISO 13849-1:2015, and GUV-
R 500 Kap. 2.28 „Betreiben von Trocknern für Beschichtungsstoffe““ ("Dryers for coating materials") (f
or
Germany)
Multifunction program controller RD3 with digital display
Adjustable ramp functions via program editor
Temperature safety controller class 2 acc. to DIN 12880:2007
Communication and printer interface RS 422
Fresh-air monitoring with visual and audible alarms and automatic heating shut-off
Replaceable fresh-air filter cartridge (fine-particle filter for particle sizes 1 µm to 10 µm, class M6 acc. to
EN 779:2012)
FKM door gasket (for max. temp. 200 °C / 392°F)
2 chrome-plated racks
Rear exhaust duct, diameter 100 mm / 3.94 inch
Options / accessories
Access ports with various diameters, with silicone plug
Rack, chrome-plated or stainless steel
Perforated rack, stainless steel
Reinforced rack stainless steel, with 1 set rack lockings
Reinforced inner chamber with 2 reinforced racks
Lockable door
Silicone door gasket resistant to high temperatures > 200 °C / 392°F. Attention: Above 250 °C / 482°F the
gasket will age faster
Fresh air replacement filter (class M6 acc. to EN 779:2012) for particles 1µm to 10µm), with aluminum
frame
Measurement of air change rate acc. to ASTM D 5374:2005
Door flap for very quick charging for Coil-Coating/Hot Air Short Cycle
Additional measuring channel for digital specimen temperature display (with clip sensor) with analog out-
put 4-20 mA at DIN socket (DIN plug included)
Factory calibration certificate, measurement in the center
Extension of factory calibration certificate (additional value)
Measuring protocol acc. to DIN 12880:2007
Qualification folder
Neutral cleaning agent (liquid concentrate)
Stable table on wheels with castors and locking brakes
Loading ...
Loading ...
Loading ...