Binder FDL 115 Safety drying chambers, series FDL – 115-liter capacity

User Manual - Page 61

For FDL 115.

PDF File Manual, 77 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
FDL (E2.1) 02/2019 page 61/76
Remarks:
Indications on the total steam space volume: If the volume of the drying material inserted in the dryer ex-
ceeds 10% of the total steam area, it must be deducted in the calculation of the total steam volume.
Indications on the heating-up and recovery times: Values without considering the prepurge time
Procedure to determine the volumetric flow rate of exhaust air: "Calculation of the volumetric flow rate
through the exhaust duct based on the velocity of flow measured there with a propeller anemometer"
Highest permitted solvent quantity: At a drying temperature of 180 degrees Celsius and an average mo-
lecular weight of the solvent of M=100g/mole. For a different drying temperature, the highest permissible
solvent quantity must be recalculated. The calculation should be made on the basis of the "principles for
the calculation of ventilation of chamber dryers and continuous dryers" in accordance with industrial
standard EN 1539:2015, Appendix B.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C ± 3 °C / 71.6 °F ± 5.4 °F and a power supply voltage fluctuation of +/- 5%. Technical data is
determined in accordance to BINDER Factory Standard Part 1:2015 following DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.