Program table corresponding to the diagram (with setting “Step”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
60
01:30
S03
80
01:00
S04
110
03:20
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).
8.1.3 General notes on programming temperature transitions
If the tolerance limits set in the user level (chap. 10) are exceeded, the program is halted until the actual
temperature value returns to within the tolerance range. During this program interruption, the LED (3d)
flashes. Therefore, the duration of the program might be extended due to the programming of tolerances
The programming is saved even in case of a power failure or after turning off the chamber.
After program rundown the controller returns to fixed value operation showing Normal Display and equili-
brates to the temperature value previously entered in fixed value entry mode.
Before starting the program, check the set-point value entered in fixed value entry mode. After
program rundown temperature will equilibrate to this value.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
The regular door gasket made of FKM is only temperature resistant up to max. 200 °C. For
temperatures above 200 °C please use the optional door gasket made of high temperature
resistant silicon.
8.2 Set-point entry for program operation
From Normal Display, press down button X/W for 5 sec to access the program editor. Then enter the set-
points one after the other in all program sections of a selected program.
You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in
the user level, chap. 10).
In order to avoid incorrect programming, we recommend entering the values of the program course into a
table (template in chap. 8.3).
Example of program table (with default setting “Ramp”):
Section
SEC
Temperature set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
60
01:30
S03
90
01:00
S04
90
03:20
S05
110
00:01
The values of the program table can now be entered to the RD3 program controller.