Binder FDL 115 Safety drying chambers, series FDL – 115-liter capacity

User Manual - Page 36

For FDL 115.

PDF File Manual, 77 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
FDL (E2.1) 02/2019 page 36/76
Then define the settings for the program course. Two parameters can be set:
Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of 1
minute, and its maximum value is 99.59 (99 hs 59 min). If the value is 00.00, the program will start im-
mediately.
Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be
entered. If the program is not going to be repeated, enter the value “0”. For infinite repeats enter the
value “–1”. The program is repeated as a whole; it is not possible to repeat individual sections.
In the last step start the selected program. These steps must be carried out subsequently.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
Step 1 Program selection (only with program type “2 programs” set):
Normal Display
Press program key
Display 1 shows
e.g. 1
(actual selection of the program)
Display 2 shows
SEL.PRG.
(select program 1 or 2)
Enter program number 1 or 2 using arrow
keys
Value is shown in display 1.
Next step entry of program course settings
Press program key
Display 1 shows
e.g. 00.00
(entered delay time hh.mm)
Display 2 shows
RUN TIME
(enter delay time of program start)
Set delay time in hh.mm using arrow keys
Value is shown in display 1.
Press program key
Display 1 shows
e.g. -1
(actual selection of the number of program cycles)
Display 2 shows
REPEAT
(enter number of program cycles)
Select number of cycles -1, 0, 1 etc. using
arrow keys
Value is shown in display 1.
Last step program start:
Press program key
Display 1 shows
e.g. 1
(selected program)
Display 2 shows
RUN PRG.
(Question: start selected program?)
Press program key
Display 1 shows
e.g. 25.5
C
actual temperature value
Display 2 shows
P01:S01 00:29:39
(time running backwards)
(actual program P01, actual section S01, and remaining
time of program section S01)
Program is running. The green LED (3d) lights up.
In addition to the green LED (3d) indicating a running program, the LED (3a) is lit if the heating is active, or
no LED if the actual temperature equals the set-point.
Loading ...
Loading ...
Loading ...