Binder FDL 115 Safety drying chambers, series FDL – 115-liter capacity

User Manual - Page 60

For FDL 115.

PDF File Manual, 77 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
FDL (E2.1) 02/2019 page 60/76
Doors
Number of doors
1
Interior dimensions
Width
mm / inch
600 / 23.62
Height
mm / inch
435 / 17.13
Depth
mm / inch
435 / 17.13
Interior volume
l / cu.ft
115 / 4.1
Total steam space volume
l / cu.ft.
156 / 5.51
Racks
Quantity of racks (regular)
2
Quantity of racks (max.)
5
Maximum load per rack
Kg /lbs
30 / 66
Maximum permitted total load
Kg /lbs
60 / 132
Weight
Weight of the chamber (empty)
Kg /lbs
88 / 194
Temperature data
Temperature range, 10 °C above ambient up to
°C / °F
300 / 572
Maximum temperature of the heating surface
°C / °F
750 / 1382
Maximum permitted oven temperature
°C / °F
300 / 572
Temperature fluctuation
at 150 °C / 302 °F
± K
0.8
Temperature uniformity (variation)
at 50 °C / 122 °F
± K
0.5
at 150 °C / 302 °F
± K
2.5
at 300 °C / 572 °F
± K
5.8
Heating-up time
to 50 °C / 122 °F
min
15
to 150 °C / 302 °F
min
20
to 300 °C / 572 °F
min
60
Recovery time after door was opened
for 30 sec.
to 50 °C / 122 °F
min
8
to 150 °C / 302 °F
min
12
to 300 °C / 572 °F
min
20
Ventilation data
Air circulation
approx. x/min.
20
Air change acc. to EN 1539:2015 at 50 °C
approx. x/min.
2.5
Volumetric flow rate of exhaust air acc. to EN 1539:2015 at
50 °C
approx. l/min
400
approx. m
3
/h
24.0
Solvent data
Highest permitted solvent quantity acc. to EN 1539:2015
(at T-180 °C, M-100g/mol, U-40g/m3, K=0,5)
g 6.65
Electrical data
Housing protection acc. to EN 60529
IP
33
Nominal voltage (+5 %)
at 50 Hz power frequency
V
230
at 60 Hz power frequency
V
230
Current type
1N~
Nominal current
Amp
13.0
Nominal power
kW
2.90
Power plug
shock proof plug
Over-voltage category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Environment-specific data
Noise level (mean value)
dB(A)
57
Energy consumption
at 150 °C / 302°F
Wh/h
1098
Loading ...