FDL (E2.1) 02/2019 page 38/76
Selection of controller menu language (Language)
Select German, English, or French.
Counter of operating hours (Oper.hs)
Information about the number of operating hours currently reached or since the last reset. (no setting,
display only).
Max. number of operating hours (Op.limit)
Enter a limit number of operating hours, i.e., the maximum number of operating hours that can be run.
Maximum setting: 9999. Reaching the limit has no effect.
Reset operating hours (Op.back)
Reset operating hours to zero.
Interface protocol (Protocol)
Modbus: The chamber interface can be used as a communication interface to connect it to a com-
puter. This serves to control the chamber by the APT-COM™ 4 Multi Management Software. It is pos-
sible to read and write the values of all parameters.
Printer: A protocol printer for data printouts can be connected to the chamber interface. The printer
regularly protocols the actual temperature value with fixed formatting and with adjustable print intervals.
In both cases an interface converter RS 422 / RS 232 is used.
Print interval (Prt.-Inv.)
Set the print interval in minutes. Function is available only if setting Printerhas been selected in the
previous menu point.
Display illumination (Disp.LED)
Select between continuous display illumination and limited illumination, which will automatically go off
300 sec after the last entry.
Program type selection (PrgSelec)
Select between entry of two programs with up to 10 sections each or of one program with up to 20 sec-
tions.
When changing from 2 programs to 1 program or vice-versa, existing programs are
deleted in the program editor.
Maximum section duration (Prg.Time)
The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs 59
min. This setting is then valid for all program sections.
When changing the maximum duration setting, pre-existing programs will be deleted in
the program editor.
Set-point programming type (Setp.sim)
Select between “Rampand “Step. With setting Stepselected, you don’t need to program the transi-
tion section in the Program Editor.
If you select setting “Step”, the controller will equilibrate only to constant temperatures;
programming ramps becomes impossible.
A change between settings “ramp” and “step” will influence all programs. Note that signifi-
cant change in time courses may arise in existing programs.