14
ENGLISH
Detector has been tested with below tables X-ray condition. This table is only for reference. The legally certified
radiologist expert should control X-ray dose.
Adult
SID (Inch / Cm) Tube Voltage (KV) Tube Current (mA) Tube Current x Time
(mAs)
Chest P-A 72 inch / 182.8cm 120 KV 250 mA 2.5 mAs
C-spine LAT 72 inch / 182.8cm 70 KV 200 mA 10 mAs
L-spine A-P 40 inch / 101.6cm 75 KV 320 mA 20 mAs
Abdomen A-P 40 inch / 101.6cm 75 KV 320 mA 10 mAs
Pelvic A-P 40 inch / 101.6cm 70 KV 320 mA 16 mAs
Wrist A-P 40 inch / 101.6cm 55 KV 100 mA 2.5 mAs
Elbow A-P 40 inch / 101.6cm 55 KV 100 mA 3.2 mAs
Shoulder AP 40 inch / 101.6cm 70 KV 200 mA 6.4 mAs
Foot A-P 40 inch / 101.6cm 58 KV 100 mA 2.5 mAs
Ankle A-P 40 inch / 101.6cm 59 KV 100 mA 3.2 mAs
Knee A-P 40 inch / 101.6cm 58 KV 100 mA 6.4 mAs
Regarding paediatric dosage, it should be much less than for an adult. The certified radiologist should pay
special attention to paediatric X-ray dosage levels.
GRID
Item Recommended Specification
SID 100 cm / 130 cm / 150 cm / 180 cm
(39.3 inch / 51.1 inch / 59 inch / 70.8 inch)
Size 460 x 460 mm
(18.1 x 18.1 inch)
Ratio 10 : 1
Frequency 215 Line / Inch
Inter Spacer AL
Battery
Item Specification
Model
Size (Width x Height x Depth) 204.6 x 110.5 x 7.8 mm
(8.0 x 4.3 x 0.3 inch)
Weight Typ. 0.24 kg (0.5 lbs)
Output Nominal voltage Typ. 7.5 V
Operation Temp 10 °C (50 °F) ~ 35 °C (95 °F)
Charging time 4 hours (standard) when charging with the Detector.
3 hours (standard) when charging two batteries with the charger.
Capacity Typ. 4000 mAh, Min. 3850 mAh
Battery performance Typ. 240 shots/6 hours
Min. 160 shots/4 hours (Cycle time 90 Sec, with Full charged battery)