Rigol DHO1202 oscilloscope

DHO1000 series user manual - Page 30

For DHO1202.

PDF File Manual, 269 pages, Read Online | Download pdf file

DHO1202 photo
Loading ...
Loading ...
Loading ...
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown
in
Figure 4.7
.
3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal
Output Terminal" of the probe, as shown in
Figure 4.7
.
Figure 4.7 Using the Compensation Signal
4. Set the probe ratio based on the attenuation of the probe, and then click or tap
> Auto.
5. Observe the waveform on the display. In normal condition, you should see a
square waveform similar to the waveform shown in the figure below.
Figure 4.8 Square Waveform Signal
6. Test the other channels in the same way. If you see the waveform, but the square
wave is not shaped correctly as shown above, perform the procedure described in
Probe Compensation
. If you do not see the waveform, perform these steps again.
WARNING
To avoid electric shock when using the probe, please make sure that the insulated wire of
the probe is in good condition. Do not touch the metallic part of the probe when the
probe is connected to high voltage source.
Quick Start
DHO1000 User Guide
16
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.