Figure 16.2 Minimized "Play" Menu
Before playing the waveforms, set the following parameters.
1. Play Mode
You can play the waveforms in single mode ( ) or cycle mode ( ).
- : plays from the start frame to the end frame, and then stops automatically.
-
: plays from the start frame to the end frame, then such playback operation
is repeated until you stop it manually.
2. Play Sequence
You can play back the waveforms clockwise ( ) or counterclockwise ( ).
- : plays from the start frame to the end frame.
- : plays from the end frame to the start frame.
3. Interval
The playback interval indicates the time interval between the frames during the
playing process.
Click or tap the input field of Interval, and then use the pop-up numeric keypad to
set the timeout value. You can also use the corresponding multipurpose knob to
set the value. The available range is from 1 ms to 1 s.
4. Start Frame
Waveform Recording and Playing
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DHO1000 User Guide
203