SDS6000A Series Digital Oscilloscope User Manual
int.siglent.com 91
display. This is identical to the "Auto" memory management mode.
14.2.4 Roll Mode
Press the Roll button on the front panel to enter roll mode. In this mode, the
waveform moves across the screen from right to left, similar to a strip chart
recorder. The horizontal delay control of the waveform will be disabled when
roll mode is active. It only operates at timebase values of 50 ms/div and above.
If you would like to stop the display in Roll mode, press the Run/Stop button.
To clear the display and restart an acquisition in Roll mode, press the Run/Stop
button again.
Zoom is only supported after Stop in Roll mode.
14.2.5 Sequence
Touch Acquire>Sequence to recall sequence dialog box.
A. Turn on or off sequence
B. Set the segment
Sequence mode is a fast acquisition mode, which divides the memory depth