int.siglent.com 299
A. Set the input low current value. This is the
expected low current value before or after the
load change
B. Set the input high current value. This is the
expected high current value before or after
the load change
C. Return to the previous menu
Note: The low and high current values are used to calculate the trigger level
and adjust the oscilloscope vertical scale. After the load changes, the current
value changes from low to high (or from high to low), and the oscilloscope
triggers and performs to measure the transient response stability time.
Connection Guide
Touch Signal>Connection Guide to recall the connection guide of transient
response, as shown in the figure below. Please follow the instructions in this
figure for connection. Touch the icon on the top right of the guide to close.
PSRR
The power supply rejection ratio (PSRR) test is used to determine how the
regulator suppresses ripple noise in different frequency ranges. The
Loading ...