SDS6000A Series Digital Oscilloscope User Manual
int.siglent.com 147
If you set the trigger condition to 7 address & data or
10 address & data:
Address can be selected in the hexadecimal
range of 0x00 to 0x7F (7-bit) or 0x3FF (10-bit). If
the address is selected as "0xXX (7-bit address)"
or "0xXXX (10-bit address)", the address is
ignored. It will always trigger on the Ack bit
followed by the address.
Data1 and Data2 can be selected in the
hexadecimal range of 0x00 to 0xFF. If the data is
selected as "0xXX", the data is ignored. It will
always trigger on the Ack bit followed by the
address.
R/W Bit can be specified as Write, Read or Don't
Care.
Data Length The data length range is 1 to 12 bytes. When the SDA data
length is equal to the value of Date Length and address length is the same as
the set value, the oscilloscope will be triggered.
Touch the Address Length to select "7-bits" or "10-bits" to match the
address of the input signal.
Touch the Data Length and turn the universal knob or virtual keypad to set
the data length to match the data length of the input signal.