that meets the trigger condition.
b) Select the "Select" mode, then set the specified line and the
specified field to trigger the signal. Assuming that the "Field" is set
to 8, you can select any field from 1 to 8, and each field can choose
any line from 1 to 100.
16.5.6 Window Trigger
Window trigger is similar to the edge trigger, except that it provides an upper
and a lower trigger level. The instrument triggers when the input signal passes
through the upper level or the lower level.
There are two kinds of window types: Absolute and Relative. They have
different trigger level adjustment methods. Under the Absolute window type, the
lower and the upper trigger levels can be adjusted separately. The relative
window provides adjustment for the Center value to set the window center and
the Delta value to set the window range. In this mode, the lower and the upper
trigger levels always move together.
If the lower and the upper trigger levels are both within the waveform
amplitude range, the oscilloscope will trigger on both rising and falling