SDS6000A Series Digital Oscilloscope User Manual
int.siglent.com 297
of its maximum amplitude
Turn Off -- Determines the time it has taken for the power supply to be turned
off to fall to a certain percentage of its maximum output voltage. The turning off
time is between T2 and T1, where:
T1 = When the input voltage finally falls to a certain percentage (usually 10%)
of its maximum amplitude
T2 = When the output DC voltage finally falls to a certain percentage (usually
10%) of its maximum amplitude
Connection Guide
Touch Signal>Connection Guide to recall the connection guide of turn on/turn
off, as shown in the figure below. Please follow the instructions in this figure for
connection. Touch the icon on the top right of the guide to close.
Transient Response
Transient response analysis can determine the response speed of the output
voltage of the power supply to the change of the output load. This time starts
from the first time that the output voltage exits the stable band and ends at the