FLUKE 284-U Arbitrary Waveform Generator and Manager, 40 MS/s, 4 Channel

User Manual for the Fluke 284-U Arbitrary Waveform Generator and Manager, 40 MS/s, 4 Channel - Page 131

For 284-U.

PDF File Manual, 160 pages, Read Online | Download pdf file

284-U photo
Loading ...
Loading ...
Loading ...
281, 282, 284
Users Manual
16-20
*
ESR? Returns the value in the Standard Event Status
Register in <nr1>
numeric format. The
register is then cleared. The syntax of the
response is <nr1><rmt>.
*
IDN? Returns the instrument identification. The
exact response is determ
ined by the
instrument configuration and is of the form
<NAME>,<model>,0,<version><rmt>
where <NAME> is the manufacturer’s name,
<MODEL> defines the type of instrument and
<VERSION> is the revision level of the
software installed.
*
IST? Returns ist
local message as defined by
IEEE Std. 488.2. The syntax of the response
is 0<rmt>, if the local message is false or
1<rmt>, if the local message is true.
*
OPC Sets the Operation Complete bit (bit 0) in the
Standard Event Status Register. This will
happen immediately
the command is executed
because of the sequential nature of all
operations.
*
OPC? Query Operation Complete status. The syntax
of the response is
1<rmt>. The response
will be available immediately the command is
executed because of the sequential nature of
all operations.
*
PRE <nrf> Set the Parallel Poll Enable Register to the
value <nrf>
.
*
PRE? Returns the value in the Parallel Poll Enable
Register in
<nr1> numeric format. The
syntax of the response is <nr1><rmt>.
*
SRE <nrf> Set the Service Request Enable Register to
<nrf>.
*
SRE? Returns the value of the Service Request
Enable Register in <nr1>
numeric format.
The Syntax of the response is <nr1><rmt>.
*
STB? Returns the value of the Status Byte Register
in <nr1>
numeric format. The syntax of
the response is <nr1><rmt>.
*
WAI Wait for Operation Complete true. As all
co
mmands are completely executed before the
next is started this command takes no
additional action.
*
TST? The generator has no self-test capability and
the response is alway
s 0<rmt>.
EER? Query and clear execution error number
register. The response format is nr1<
rmt>.