16.4.3 Time Base Calibration
Time base calibration is used to adjust the time base inside the device to its most accurate state.
This calibration requires the use of an external 10 MHz standard frequency source. The
connection relationship is shown in the figure below. Connect the 10 MHz output of the frequency
source to the frequency meter input of the device:
Frequency
Standard
SDG7000A
10 MHz
Reference
Figure 16-2 time base calibration environmen
After configuring the calibration environment according to the above figure, ensure that the
standard frequency source and the heat engine time of the device exceed the specified minimum
time, execute Utility > Test/Cal > TimeBase Cal to enter the relevant setting page.
Frequency values measured by frequency meters
The DAC code value corresponding to the voltage control voltage of the current
time base oscillator, where the code value can be entered for manual correction
Perform automatic calibration
Restore the default configuration
Save the correction results
Cancel this calibration