UF V (E3) 10/2020 Page 97/119
Fault description
Possible cause
Required measures
Controller
No chamber function
(dark display).
Power failure. Display mode
“Standby” active.
With option “battery-buffered alarm
system”: Press any key on the con-
troller display.
Main power switch (4) is off.
Turn on the main power switch
(4).
Menu functions not available.
Menu functions not available with
current authorization level.
Log in with the required higher au-
thorization.
No access to controller
Password incorrect.
Contact BINDER service.
Acknowledging the alarm does
not cancel the alarm state.
Cause of alarm persists.
Remove cause of alarm. If the
alarm state continues, contact
BINDER service.