UF V (E3) 10/2020 Page 35/119
4.4 Connections of cooling water for chambers with water cooling
The water cooling reduces the heat, which is emitted during cooling operation to the ambient air.
An enclosure inside the chamber contains the connection kit for the cooling water inlet and outlet.
„OUT“ „IN“
Figure 19: Connections of cooling water on the chamber rear (chambers with water cooling), example UF
V 700
“IN” connection for cooling water inlet with external thread 3/4‘‘ and internal thread 3/8‘‘
“OUT” connection for cooling water outlet with external thread 3/4‘‘ and internal thread 3/8‘‘
4.4.1 Connection of cooling water outlet for water cooling
Fasten the 1/2‘‘ cooling hose to the connection of cooling water outlet “OUT” on the chamber rear. Observe
the following points:
You can use a part of the supplied water hose for the cooling water outlet. In case another hose is used,
it must be permanently resistant against max. 50 °C / 122 °F and pressure-resistant up to 10 bar.
Put the hose on the hose nozzle with screwing and secure it with one of the four supplied hose clamps.
Connect the hose nozzle to the connection OUTand screw on the union nut.
For the hose connection to the domestic water connection, we recommend to also use the supplied
hose nozzle with screwing and secure it with one of the four supplied hose clamps.
Before turning on the chamber, check the connection for leaks.
The temperature of the effluent cooling water is 27 °C up to 29 °C on account of the chamber’s construction.