22. Data monitoring and recording
22.1 Ethernet interface
The chamber is regularly equipped with an Ethernet interface (8) that can connect the BINDER APT-COM™
4 Multi Management Software (option). The MAC Address is indicated in the “Ethernet” controller menu
(chap. 15.1.1).
22.2 APT-COM™ 4 Multi Management Software (option)
You can connect the BINDER APT-COM™ 4 Multi Management Software via the chamber’s Ethernet in-
terface (8).
The APT-COM™ System permits the networking of up to 100 chambers and connection to a PC for con-
trolling and programming, as well as recording and representing temperature data. The actual temperature
value is given at adjustable intervals. Programming can be performed graphically via PC. For further infor-
mation please refer to the APT-COM4 operating manual.
22.3 Analog output for temperature (option)
With this option the freezer is equipped with an analog output 4-20 mA for temperature. This output permits
transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (10) in the rear connection panel as follows:
ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature
PIN 2: Temperature +
Temperature range: +40 °C / 104°F to -100 °C / -148°F
A suitable DIN plug is enclosed.
Figure 24: Pin allocation of DIN socket (10) for option analog output
22.4 Data logger kit (option)
The BINDER Data Logger Kit offers an independent long-term measuring system for temperature. It is
equipped with a keyboard and a large LCD display, alarm functions and a real-time function. Measurement
data are recorded in the Data Logger and can be read out after the measurement via the RS232 interface
of the Data Logger. It offers a programmable measuring interval and permits storing up to 64000 measuring
values. Reading out is done with the Data Logger evaluation software. You can give out a combined alarm
and status protocol directly to a serial printer.
Data Logger Kit T 220: Temperature range -90 °C / -130 °F up to +220 °C / 428 °F
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the