Binder UF V 700 UF V 500-liter Ultra Low Temperature Freezer

User Manual - Page 49

For UF V 700.

PDF File Manual, 119 pages, Read Online | Download pdf file

UF V 700 photo
Loading ...
Loading ...
Loading ...
UF V (E3) 10/2020 Page 49/119
Or
Setting the safety controller value with “Offset” safety controller mode
The current value flashes. Enter the desired offset value with the arrow
buttons.
Confirm the entry with the OK button.
Offset
With the Back button you can go back to the “Safety controller” submenu and, repeatedly pressing it, to
Normal display.
Regularly check the settings of the safety controller mode and value.
Set the safety controller value by approx. 15 °C above the desired temperature set-point.
11.3 Message and measures in the state of alarm
The state of alarm is indicated visually in Normal display. If the buzzer is enabled (chap. 14.3) there is an
additional audible alert. A text message indicates the alarm cause. The “collective alarm” icon is lit. If the
audible arm is activated, the buzzer sounds. Press the OK button to mute the buzzer.
The alarm message “Safety controller” and the “Collective alarm” icon are displayed on the controller until
you press the OK button on the controller and the inner chamber temperature has cooled down below the
safety controller value.
If the inner chamber temperature has already cooled down below the safety controller value when press-
ing the OK button, the alarm message Safety controllerand the Collective alarmicon are reset
together with the buzzer.
If the state of alarm is still active when pressing the OK button, i.e. the inner chamber temperature is
still above the safety controller value, first only the buzzer is reset. The alarm message Safety controller
and the Collective alarmicon will disappear as soon as the inner chamber temperature falls below the
safety controller value.
Normal display showing safety controller alarm (sample value)
Safety controller
Note:
When the safety controller had been activated you should disconnect the chamber from the power supply
and have an expert examine and rectify the cause of the fault.
11.4 Function check
Check the safety controller at appropriate intervals for its functionality. It is recommended that the author-
ized operating personnel should perform such a check, e.g., before starting a longer work procedure.
Loading ...
Loading ...
Loading ...